Skip to content

SCPI Reference

Val edited this page Nov 26, 2026 · 11 revisions

Commands work over USBTMC and the CDC serial port (USB and Remote Mode). This page lists the most useful commands.

Syntax

  • Short and long forms (TRIG:EDGE:LEV = TRIGger:EDGE:LEVel), case-insensitive, ; between commands, LF or CRLF.
  • Units: Hz/kHz/MHz, V/mV/uV, s/ms/us/ns, %.
  • Errors go to SYST:ERR?: -113 unknown command, -222/-224 bad value, -221 not possible in the current state, -230 no data yet. A failed query returns NAN.
  • An unsupported feature returns an error.

Common

Command
*IDN? OpenOSC,<profile>,<serial>,<version>; the profile name is a compatibility alias
*CLS, SYST:ERR?, *ESR?, *STB? Status and error queue
*OPC?, *WAI Wait for multimeter commands and single captures
SYST:MODE SCOPe|DMM|GENerator Select the Oscilloscope, Multimeter or Generator screen directly
SYST:REM ON|OFF Remote mode
SYST:KEY <key>[,<ms>] Press a panel key. For example, SYST:KEY F1,1200 holds F1 for 1.2 s. *OPC? waits for release and handling
SYST:CLOC:MODE 200|280|360 Performance Profiles
SYST:REC Restart into Recovery
SYST:CAL:FACT?, SYST:CAL:FACT:FORM? The calibration record as read at boot, and its format
DISP:DATA? (HCOP:DATA?) The screen as a BMP block (320×240)
SYST:STALl? Main-loop passes of 50 ms or more and Auto frame gaps of 1 s or more, with what ran then
SYST:BOOT? Milliseconds of each boot stage

Oscilloscope

Command
RUN, STOP, SING, TFORce Capture control
TRIG:STAT? RUN, WAIT, STOP; TD for a triggered frame, AUTO for a free-running one
CHAN<n>:DISP, :COUP, :PROB, :SCAL, :OFFS Channel settings. Probe ratios: 1, 10, 50, 100, 500, 1000, 10000. For example, CHAN1:PROB 10000; CHAN1:PROB? reads the ratio
TIM:SCAL, TIM:OFFS Timebase (rounded to the nearest step) and horizontal position
ACQ:MDEP <points>, ACQ:MDEP? Request a depth per enabled channel. OpenOSC picks the smallest matching total budget. Query the planned sample count per channel. For the actual captured count, read WAV:PRE? with a full record window and point reduction off
ACQ:SRAT?, ACQ:ROLL Planned sample rate and Roll
ACQ:INTerleave ON|OFF|AUTO, ACQ:INTerleave? The Interleave CH1 option, saved like the menu choice. Queries return 1, 0 or AUTO. The automatic rule depends on the model (Performance Profiles)
ACQ:INTerleave:MATCh? ZT703S: 1 when a gain and timing profile is in use for the configured capture (Calibration)
ACQ:INTerleave:STATe? <active>,<blockers>: 1 when interleave is in force; blocker bits 1 board, 2 run mode, 4 timebase, 8 channels, 16 Roll, 32 Auto Setup or trial, 64 generator DAC, 128 option off. The rate is in ACQ:SRAT?
TRIG:SWE AUTO|NORM|SING, TRIG:EDGE:SOUR, :SLOP, :LEV, :NREJ, TRIG:POS Trigger
AUToscale Start Auto Setup; works with either physical AUTO key mapping
CHAN<n>:FCNT?, MEAS:FCNT?, MEAS:FCNT:QUAL? Hardware frequency counter and its quality
WAV:SOUR, WAV:MODE RAW, WAV:FORM BYTE, WAV:STAR, WAV:STOP, WAV:PRE?, WAV:DATA? Read waveforms (up to 524,288 points per channel)
WAV:POINts <n> Return at most n points: every Nth point of START..STOP; the preamble matches
WAV:SNAP, WAV:SEQ?, ACQ:FRAMe? Frame handling: release the held frame, frame numbers

Since 0.21.19, qualified trigger commands set the trigger type, levels, polarity or slope and time condition. Trigger SCPI commands

Desktop waveforms and measurements

These extensions are available from 0.21.9. Both USBTMC and CDC support them. The command reference defines the binary blocks, preambles, measurement fields and timeout behavior. Desktop commands

Command Result
WAV:FRAM? Enabled calibrated channels from one acquisition, in one block
WAV:POIN:MODE STR or PEAK Choose stride sampling (default) or min/max pairs when WAV:POIN requests point reduction
WAV:ENV? [buckets] Min/max pairs over the full acquisition, with 1500 buckets by default
WAV:MEAS? SCR or REC Measurements for the selected sample window or the full record on the selected channel

In RUN, WAV:FRAM? and WAV:ENV? wait for a fresh capture. Set the host read timeout above the configured record duration plus one second. In HOLD and after a completed Single, they return the held record.

In the 0.22.0 pre-release, with Short window f/T → From the counter selected, the final frequency_source field of WAV:MEAS? may be CNT or CNT?. Both indicate frequency and period from the hardware counter. CNT? is the approximate case. The other sources are NONE, EDGE and SPEC. The Trace f (~) display option leaves CHAN<n>:FCNT? and MEAS:FCNT? reporting the hardware counter. Measurement fields

Files and idle settings

Command Result
MMEM:CAT? "FOLDER"[,start] List finished files and their sizes, with pagination
MMEM:DATA? "FOLDER/FILENAME" Download a finished file as a binary block
MMEM:DEL "FOLDER/FILENAME" Delete a finished file. A file in use returns -221
SYST:POW:TIM[?] Read or set auto power-off in minutes
SYST:DISP:TIM[?] Read or set backlight timeout in seconds
SYST:DISP:DIM[?] Read or set dimmed backlight brightness in percent

Files are available in SCREENSHOTS, LOGS, CAPTURES, DECODE and FRA under /openosc. In the 0.22.0 pre-release, list CAPTURES to find .CSV.GZ and .RAW.GZ archives and download their gzip bytes using MMEM:DATA?. File commands and binary format

Protocol decoder

Command
DECode:STATe?, DECode:PROTocol? Is the decoder on, and which protocol: OFF, UART, I2C, SPI, LIN, CAN, RS232, RS485 or DMX512
DECode:RESult? Summary of the latest result: status, events, lost events, sample rate …
DECode:DATA? The decoded events as CSV
DECode:SAVE, DECode:SAVE? Save the CSV on the device; path of the last file

Change decoder settings on the device (Protocol Decoders). The ac_coupled result status asks you to set the RS-485 or DMX512 input to DC. DMX512 CSV v3 includes data slot numbers and BREAK and MAB duration. Serial decoder results

Multimeter

Command
CONF:VOLT:DC|AC, CONF:CURR:DC|AC, CONF:MVOL:…, CONF:MCUR:…, CONF:RES, CONF:CAP, CONF:CONT, CONF:DIOD, CONF:TEMP Select the function; CONF:TEMP selects Type K
FUNC? Current function (TEMP for Type K)
SYST:DMM:TEMP:CJUN <°C>|NONE, SYST:DMM:TEMP:CJUN? Reads the stored manual Type K junction value. A write selects Manual with the supplied value (−50.0 … 100.0 °C in 0.1 °C steps); NONE clears it and returns to DC mV. Select MCU at startup or MCU live from the device menu
SYST:DMM:TEMP:UNIT C|F, SYST:DMM:TEMP:UNIT? Display unit; readings over SCPI are always in °C
MEAS?, FETCH? Latest reading (respects HOLD and REL)
READ? Next new reading (waits up to 2 s)
AUTO, RANGE <n> Auto range, manual range index
SYST:DMM:RANG:LIST? Full-scale values of the ranges of the current function
SYST:DMM:RES? Reading and its metadata in one line
SYST:DMM:HOLD ON|OFF, SYST:DMM:HOLD:MODE MAN|DEL|STAB|AUTO, …:DEL, …:STAB, …:BAND, …:REAR, …:COUN, …:ARM, …:STAT? HOLD and HOLD modes
SYST:DMM:REL ON|OFF Relative measurement
SYST:DMM:STAT?, SYST:DMM:STAT:RES Statistics, reset
SYST:DMM:PACK? The newest raw packet from the meter chip, in hex

SYST:DMM:RES? fields: function, value, status, secondary value, secondary quantity, range index, range mode (1 auto, 0 manual, −1 unknown), HOLD, REL, sequence, timestamp in µs. Status 14 means the meter showed no number. The DC charge counter is a local display feature.

Signal generator

Command
SOUR:FUNC SIN|SQU|TRI|HALF|POS|RAMP Waveform
SOUR:FREQ <f> Frequency
SOUR:PULS:DCYC <%> Duty (square)
OUTP ON|OFF Output
OUTP:STAR OFF|REST Output at power-on
SOUR:LEV:PROF, SOUR:LEV:LOW, SOUR:LEV:HIGH, SOUR:LEV:PATH?, SOUR:SQU:DRIV ZT703S: level profile, DAC codes, output path, square drive
SOUR:VOLT <V> ZT706S: amplitude; units such as mV are accepted

On the other model, model-specific commands return -241 Hardware missing.

Example

Read one DC voltage (multimeter screen active):

*IDN?
CONF:VOLT:DC
READ?

Take a single capture and read CH1 (oscilloscope screen active):

TIM:SCAL 1e-3
SING
*OPC?
WAV:SOUR CHAN1
WAV:MODE RAW
WAV:FORM BYTE
WAV:PRE?
WAV:DATA?

SYST:MODE between oscilloscope and multimeter re-connects USB, so reopen the connection after it before sending the next command.