Add this suggestion to a batch that can be applied as a single commit.This suggestion is invalid because no changes were made to the code.Suggestions cannot be applied while the pull request is closed.Suggestions cannot be applied while viewing a subset of changes.Only one suggestion per line can be applied in a batch.Add this suggestion to a batch that can be applied as a single commit.Applying suggestions on deleted lines is not supported.You must change the existing code in this line in order to create a valid suggestion.Outdated suggestions cannot be applied.This suggestion has been applied or marked resolved.Suggestions cannot be applied from pending reviews.Suggestions cannot be applied on multi-line comments.Suggestions cannot be applied while the pull request is queued to merge.Suggestion cannot be applied right now. Please check back later.
The NA500 BNC input was configured in mode 1 (events only), so incoming pulses did not enable FPGA EEG blanking. Configure a fixed 1000 µs blanking width before selecting mode 2 during stream setup, including when joining an already-running amplifier and after reinitialization. Setup fails if either command is rejected.
The mock continues reporting BNC events in mode 2. This also adds repeatable hardware capture/analysis utilities, an example DHCP-only configuration for a dedicated amplifier link, and a hardware report with the comparison plot and quantitative results. Large raw recordings remain local and are ignored by Git.
Validation
cmake --build build -j 6).git diff --checkpassed.All comparison blocks had contiguous counters and 250 µs timestamp increments; none of the analyzed event windows showed EEG saturation. Mode 2 produced a repeatable EEG switching transient whose later portion moved later with increased blanking width, while mode 1 stayed near background noise. This supports a width-dependent hardware response and preserved events. A driven-input test is still needed to verify suppression of a known signal or TMS artifact; this open-input experiment does not establish suppression or exact analog gate timing.
The full method, plot, measurement summary, and reproduction commands are in the hardware report.