Original file line number Diff line number Diff line change
Expand Up @@ -109,7 +109,7 @@ Bypasses the FPGA filter and samples directly at the requested rate. This provid
- Reduced bandwidth (~1/4 of sample rate, e.g., 250 Hz at 1000 Hz sample rate)
- Optimized for EEG-TMS and real-time BCI applications

Available native rates: 500, 1000, 2000, 4000, 8000 Hz (NA400); 250, 500, 1000, 2000, 4000, 8000, 16000 Hz (NA500)
Available native rates: 500, 1000, 2000, 4000, 8000 Hz (NA400); 250, 500, 1000, 2000, 4000 Hz (NA500)

Use `--fast-recovery` (or a "Low Latency" dropdown entry) to select native mode for rates where the model offers both.

Expand All @@ -131,14 +131,31 @@ The NA500 rows are zero deliberately: Net Station 6 applies no anti-alias alignm

## Net Amps 500 (GES 500) Support

NA500 support was verified against real hardware (firmware 0.4.9). Differences from the NA400 that this app handles automatically:
NA500 support was verified against real hardware (firmware 0.4.9). The amplifier's protocol is documented, as observed on the wire, in [docs/NA500_AmpServer.md](docs/NA500_AmpServer.md).

**Finding the amplifier**: the NA500 gets its address by DHCP and advertises itself over mDNS (service `_na500._tcp`), typically as `NA500-256ch-<serial>.local`. Pass that name instead of an IP, e.g. `--address NA500-256ch-0001.local`. This works out of the box on macOS and recent Windows 10/11; Linux needs Avahi with `nss-mdns`. To list amplifiers on the network, run `dns-sd -B _na500._tcp` (macOS/Windows with Bonjour) or `avahi-browse -r _na500._tcp` (Linux). With no DHCP server on the network the amplifier has no IPv4 address, and this app cannot reach it because its AmpServer is IPv4-only. When the amplifier is cabled directly to this computer (or to a switch with no DHCP server), give the computer's adapter a static address and serve the amplifier a lease with [dnsmasq](https://thekelleys.org.uk/dnsmasq/doc.html) (`brew install dnsmasq` / `apt install dnsmasq`). For example, with the adapter `en10` set to 10.10.10.61/24:

```bash
sudo dnsmasq -d --port=0 --interface=en10 --bind-interfaces \
--dhcp-range=10.10.10.62,10.10.10.62,255.255.255.0,12h
```

`--port=0` disables dnsmasq's DNS server, so it only hands out the one address; `-d` keeps it in the foreground (Ctrl+C stops it). The amplifier picks the lease up within a minute, without a restart, and it must be running whenever the amplifier powers up.

Differences from the NA400 that this app handles automatically:

- **Scaling**: 0.044703 µV/bit (±375 mV over 24 bits, matching the GES 500 datasheet and Net Station 6).
- **Rates**: decimated up to 2000 Hz; native from 250 Hz up to 16000 Hz. No duplicate packets are sent at low rates (the NA400 duplicates samples below 1000 Hz).
- **Rates**: decimated up to 2000 Hz; native from 250 Hz up to 4000 Hz. No duplicate packets are sent at low rates (the NA400 duplicates samples below 1000 Hz). 8000 and 16000 Hz are not offered: the amplifier accepts them but delivers at most about 7995 samples/s, and the missing samples leave no gap in its packet counter (about 4 samples/s lost at 8000 Hz, half at 16000 Hz).
- **Initialization**: the NA500 silently ignores commands sent while it is powering up (~10 s from cold), so the app waits for the power-on notification before configuring it.
- **Data connection handling**: the NA500's AmpServer stops serving data to new listeners once any listener disconnects (until the amp restarts), so the app maintains one persistent data connection per session and never issues `cmd_StopListeningToAmp`.
- **Data connection handling**: the NA500's AmpServer serves one data listener per session; once it disconnects, no new listener is served until the command connection closes and a new session starts. The app therefore maintains one persistent data connection per session and never issues `cmd_StopListeningToAmp` (which the NA500 never answers).
- **Continuous reading**: the NA500 drops samples if its data port goes unread for a few seconds, so the app does not wait for a Physio16 status notification (the NA500 sends none).
- **Stale packets**: after `cmd_Start` the NA500 first resends leftover packets from its previous stream before its counter restarts at 0; the app discards them.
- **Front end**: the NA500's EEG reads full scale until its buffered reference is switched on, every `cmd_Stop` switches it off, and switch commands sent within ~20 ms of power-on are lost; the app waits briefly after power-on and enables the buffered reference before each `cmd_Start`.
- **Physio16**: the NA500 never reports whether a Physio16 is attached, so the app always streams its 16 physio channels (PIB1-16) after the EEG. Physio gains persist on the amplifier, so the app sets them to 12, the power-on default, before each start; the physio is then scaled like the EEG (±375 mV range) and needs no realignment delay.
- **Triggers**: the app enables the NA500's DB9 digital inputs (without that they read as all 16 active) and its BNC input. Each rising edge on BNC in appears in the `_DIN` stream as bit 16 (value `0x10000`) for 1 ms, alongside the 16 DB9 bits. The EEG trails a BNC-in event by about 1.5 samples (the ADC filter's delay); no correction is applied, as in Net Station.
- **Clock**: the NA500 stamps packets with its own clock, which is not synchronized; the app sets it with `cmd_SetAmpTime` at power-on, as Net Station does.
- **Single client**: while Net Station (or any other client) is linked to the NA500, this app's connections are accepted but not serviced — and vice versa. Unlink Net Station before using this app; concurrent monitoring is not possible on the NA500 (unlike the NA400).
- **Recovery**: the NA500 emits no `ntn_AmpStarted` notification, so external-restart recovery uses data-port polling instead.
- **Recovery**: the NA500 sends `ntn_AmpStarted` only in reply to `cmd_GetStartTime`, not when streaming starts, so external-restart recovery uses data-port polling instead.
- **Robustness**: the NA500's embedded AmpServer (fw 0.4.9) can crash or stop responding after abrupt client disconnects; commands now time out after 10 s instead of hanging, and the GUI performs linking on a background thread so the UI never freezes. A crashed AmpServer requires power-cycling the amplifier.

## Timestamp Compensation
Expand All @@ -149,9 +166,9 @@ There is **no flag to toggle this** — it follows the configured mode automatic

### 1. System (pipeline) delay — always applied

Every pushed timestamp (EEG, Physio16, **and** DIN) is moved earlier by `systemdelayms` to account for the time between digitization and the sample becoming available here: device firmware + network transmission + our read path. Measured at ~5 ms in native mode using `scripts/audio_latency_test`.
Every pushed timestamp (EEG, Physio16, **and** DIN) is moved earlier by `systemdelayms` to account for the time between digitization and the sample becoming available here: device firmware + network transmission + our read path. Measured using `scripts/audio_latency_test` at ~5 ms on the NA400, and on the NA500 at ~1.0 ms plus 1.1 sample periods (5.4 ms at 250 Hz, 2.1 ms at 1000 Hz, 1.3 ms at 4000 Hz).

The default is **4.5 ms** — deliberately a little *under* the measured value. Under-compensating is the safe direction: it guarantees a sample is never back-dated to *before* the event that produced it (a response must not precede its stimulus). It is user-configurable; raise it toward the measured latency if you prefer tighter alignment and can accept that risk.
The default is **4.5 ms** (NA400 and other models) and **0.8 ms + one sample period** on the NA500 (4.8 ms at 250 Hz, 1.8 ms at 1000 Hz) — deliberately a little *under* the measured value. Under-compensating is the safe direction: it guarantees a sample is never back-dated to *before* the event that produced it (a response must not precede its stimulus). It is user-configurable; raise it toward the measured latency if you prefer tighter alignment and can accept that risk.

### 2. FPGA filter offset — decimated mode only

Expand All @@ -169,7 +186,7 @@ Both values live in the config file (and persist via File → Save Configuration
<settings>
...
<physioaligndelayms>33</physioaligndelayms> <!-- Physio16 realignment, decimated only -->
<systemdelayms>4.5</systemdelayms> <!-- pipeline-latency compensation, always -->
<systemdelayms>4.5</systemdelayms> <!-- pipeline-latency compensation, always; omit for the per-model default -->
</settings>
```

Expand Down Expand Up @@ -293,6 +310,16 @@ cmd_TurnChannelDriveSignals(N, 1) - Turn ON drive signal (reset)
cmd_TurnChannel10KOhms(N, 0) - Turn OFF 10K resistor (reset)
```

### Net Amps 500

The NA500 runs the same algorithm with these differences, handled automatically:
- **Drive level**: calibration amplitude 1023 (its DAC is 10-bit), which gives ~5000 µV p-p on every channel. The app uses Net Station's fixed 5000 µV ideal signal rather than self-calibrating, which catches a switching transient on this amplifier.
- **Oscillator last**: the NA500 latches amplitude and frequency when the oscillator turns on, so the app switches it on after setting them.
- **Reference**: measured in the NA500's COM/REF measurement mode (`cmd_SetCOMAndRefForMeasurement`), with ~1.2 s settling on each side; the EEG reads 0 during that step.
- **Leaving impedance mode**: `cmd_DefaultAcquisitionState` does nothing on the NA500, so the app restores the acquisition state explicitly.

Verified on a GES 500 with a test jig that ties all inputs together (all channels ~0 kΩ, reference ~0.5 kΩ); absolute accuracy against real electrodes has not been checked.

### Limitations and Notes

- **Tiling-based scanning**: When a tiling layout is available for the detected net (32/64/128/256 HydroCel GSN nets), channels are measured in groups (~5-6 collections per full scan) rather than one at a time — no flag required, it engages automatically. A full 256-channel scan drops from ~5 minutes to roughly a minute. Nets without a tiling layout fall back to single-channel scanning.
Expand Down Expand Up @@ -430,6 +457,14 @@ To use the mock server for development:

The mock server generates synthetic sine waves (10-50 Hz) with noise for the EEG data, so you can also verify the LSL stream in downstream applications.

The C++ mock (`MockAmpServer`, built with the project) also emulates other models. `MockAmpServer --amp-type NA500` reproduces the NA500's session-based AmpServer, including its quirks (see [docs/NA500_AmpServer.md](docs/NA500_AmpServer.md)); add `--net-code None` to emulate a unit with no net attached.

`scripts/ampserver_conformance.py` records how an AmpServer behaves on the wire and diffs two recordings. The recording of a real NA500 is in `mock/conformance/`; to check the mock against it:

> MockAmpServer --amp-type NA500 --net-code None &
> python3 scripts/ampserver_conformance.py --host 127.0.0.1 --active -o mock.json
> python3 scripts/ampserver_conformance.py --compare mock/conformance/na500_fw0.4.9.json mock.json

# Known Issues

## Net Station Acquisition Compatibility
Expand Down
9 changes: 5 additions & 4 deletions ampserver_config.cfg
Original file line number Diff line number Diff line change
Expand Up @@ -17,8 +17,9 @@
<physioaligndelayms>33</physioaligndelayms>
<!-- System/pipeline latency (ms) subtracted from every pushed timestamp
(EEG, physio, DIN) to compensate digitization->client delay (firmware +
network + read path). Measured ~5 ms via scripts/audio_latency_test;
default 4.5 deliberately under-compensates so a sample is never back-dated
before its event. User-tunable. -->
<systemdelayms>4.5</systemdelayms>
network + read path), measured via scripts/audio_latency_test: ~5 ms on
the NA400, ~1 ms + 1.1 sample periods on the NA500. When unset the app uses
a little less (4.5 ms; NA500: 0.8 ms + one sample period) so a sample is
never back-dated before its event. Setting it overrides every model. -->
<!-- <systemdelayms>4.5</systemdelayms> -->
</settings>
Loading
Loading