Scope: Software Voltage Scaling (V/Div) in Oscilloscope Mode #174
This request describes one feature or improvement, or changes that belong to the same feature.
Area
Scope
What do you want to do?
In oscilloscope mode, provide the ability to digitally zoom the vertical voltage scale (Zoom) and adjust the voltage sensitivity per division to 1 mV, 2 mV, 5 mV, and 10 mV/div. This will allow weak signals to be displayed at a larger scale, improving the accuracy of their visual analysis and measurement.
Implement this functionality similarly to the existing time-axis zoom (Time/Div), with convenient switching between voltage scales (V/Div).
changed the title [-]Software Voltage Scaling (V/Div) in Oscilloscope Mode.[/-][+]Scope: Software Voltage Scaling (V/Div) in Oscilloscope Mode[/+]on Nov 26, 2026
The DM40 has already been implemented and works very well. At least, we have the opportunity to give it a try and see for ourselves how well it performs.
Thanks for the request and the photos. I'm closing this because of hardware limits.
The ZT703S and ZT706S front end goes down to 20 mV/div, and the ADC is 8-bit. A 1, 2, 5 or 10 mV/div scale could only stretch the codes already captured at 20 mV/div. Each division would hold just a few ADC steps, so the trace would show quantization steps and front-end noise, not more detail of the signal (as TzOk83 noted above). The DM40 has a different analog front end and ADC, so its results don't carry over to these units.
For weak signals, the best options on these devices are the 20 mV/div range with DC or AC coupling as appropriate, Average acquisition for repetitive signals, and the Record readouts basis, which measures over every point of the record.
Activity
[-]Software Voltage Scaling (V/Div) in Oscilloscope Mode.[/-][+]Scope: Software Voltage Scaling (V/Div) in Oscilloscope Mode[/+]TzOk83 commentedon Nov 26, 2026
Digital zoom on the voltage axis will basically display random noise. This scope already goes to the limits (or even beyond) of its ADC.
OlegMedvedcuk commentedon Nov 26, 2026
The DM40 has already been implemented and works very well. At least, we have the opportunity to give it a try and see for ourselves how well it performs.
bve commentedon Nov 26, 2026
Thanks for the request and the photos. I'm closing this because of hardware limits.
The ZT703S and ZT706S front end goes down to 20 mV/div, and the ADC is 8-bit. A 1, 2, 5 or 10 mV/div scale could only stretch the codes already captured at 20 mV/div. Each division would hold just a few ADC steps, so the trace would show quantization steps and front-end noise, not more detail of the signal (as TzOk83 noted above). The DM40 has a different analog front end and ADC, so its results don't carry over to these units.
For weak signals, the best options on these devices are the 20 mV/div range with DC or AC coupling as appropriate, Average acquisition for repetitive signals, and the Record readouts basis, which measures over every point of the record.