Power: Backlight timeout reset on DMM value change / OSC trigger event #169
This request describes one feature or improvement, or changes that belong to the same feature.
Area
Power
What do you want to do?
In DMM mode, the backlight timer could be reset or the device could be woken from idle mode if the measured value has changed - for example, by more than 1% of the current measurement range.
In scope mode the reset action could be a trigger event.
Maybe also the auto-power off timer should be reset.
This would be really useful keeping the hands at the work piece when doing intermittent measurements during a device repair.
This kind of behaviour is for example implemented in the YR 1035+ internal resistance meter and I really miss this function in every other DMM.
changed the title [-]Backlight timeout reset on DMM value change / OSC trigger event[/-][+]Power: Backlight timeout reset on DMM value change / OSC trigger event[/+]on Nov 26, 2026
Increase delta threshold to 10% to prevent waking from static charge buildup.
Additionally wake by shorted probe (fixed window crossing: 0 ± 0.1mV)
My observation is that in DC mV the reading ends in a constant "OL" after some time anyway due to static charge buildup. AC mV seems to be a bigger problem as it is really sensitive to EMI without maxing out.
So maybe just disable it in mV completely.
Settings → Device → Power → Wake on measurement now offers Off, DMM, Scope and Both. The DMM uses a 10% same-range threshold. A genuine scope trigger or a qualifying DMM change restores the backlight and restarts the idle shutdown timer.
Activity
[-]Backlight timeout reset on DMM value change / OSC trigger event[/-][+]Power: Backlight timeout reset on DMM value change / OSC trigger event[/+]Nikonist commentedon Nov 26, 2026
The question is how to handle DC mV range...
gworky commentedon Nov 26, 2026
My observation is that in DC mV the reading ends in a constant "OL" after some time anyway due to static charge buildup. AC mV seems to be a bigger problem as it is really sensitive to EMI without maxing out.
So maybe just disable it in mV completely.
bve commentedon Nov 26, 2026
Included in the OpenOSC 0.22.0 pre-release.
Settings → Device → Power → Wake on measurement now offers Off, DMM, Scope and Both. The DMM uses a 10% same-range threshold. A genuine scope trigger or a qualifying DMM change restores the backlight and restarts the idle shutdown timer.