Scope: A second CH1 trace appears in AC coupling while CH2 is enabled #167
This report describes one problem, or problems that are clearly related.
Device
ZT706S
Firmware version
0.21.19
Area
Scope
Performance profile
200/100
Steps to reproduce
Wire the internal generator output to CH1 and set the generator to Sine, 2 MHz.
Set CH1 to AC coupling (No persistence set), 1 V/div, 100 ns/div, and leave CH2 enabled.
Power the instrument off and on again (boot with AC coupling already selected) and leave the generator running.
Watch CH1 without touching any setting while the AC coupling removes the generator's DC component.
Repeat the same steps with CH2 switched off.
Expected behaviour
One CH1 trace that moves to its final position as the coupling capacitor charges, then shows the 2 MHz sine steadily. Enabling or disabling CH2 should change nothing about CH1's trace.
Actual behaviour
While the AC coupling is still settling, CH1 is drawn as two parallel traces, one above the other and the same shape, which slowly converge into a single trace over a few seconds. After they meet the display is normal. With CH2 switched off and the same signal and CH1 settings, only one trace appears and the settling looks right, so the duplicate only shows while both channels are enabled. It is not seen with CH1 in DC coupling.
The extra Math trace now has a visible MATH label with the channel operation, placed below the frequency overlay. I checked the label with two inputs and a 2 MHz signal on ZT703S. The reported ZT706S AC-coupling sequence still needs hardware confirmation. If the extra trace appears with Math set to Off on 0.22.0, please share a new capture and the scope settings.
Activity
bve commentedon Nov 26, 2026
Included in the OpenOSC 0.22.0 pre-release.
The extra Math trace now has a visible MATH label with the channel operation, placed below the frequency overlay. I checked the label with two inputs and a 2 MHz signal on ZT703S. The reported ZT706S AC-coupling sequence still needs hardware confirmation. If the extra trace appears with Math set to Off on 0.22.0, please share a new capture and the scope settings.
eljjc commentedon Nov 26, 2026
Thanks! Yes, Math was set to On :)