Skip to content

Decoder: Decoder Save CSV Issue #164

Done

Description

@Nikonist

One problem

  • This report describes one problem, or problems that are clearly related.

Device

ZT703S

Firmware version

0.21.18

Area

Decoder

Performance profile

Not relevant / Recovery

Steps to reproduce

1)- In menu Analyze, Save CSV produce a message : "Not enough free space for this record" even if the storage has 16MB free space.
However, under Decoder, activating the 4 protocols (UART, CAN, LIN, RS232) does generate a CSV file, from the Decoder menu or from the Decoder interface, View Table. Not with the other protocols.

2)- In the Decoder interface, the F1 label is "Proto" where it should be "Protocol" (space is enough).

3)- In RS232 protocol, with no probe connected, there is a continuous flow of data (random).

Expected behaviour

1)- If Save CSV is related to any Decoder protocol, then it should be removed from the Analyze menu. If there is a reason why it's not working there, then it should be stated in the error message. Disk space is free.

2)- The protocol label should display "Protocol" in full as there is enough space (above F1 key).

3)- RS232 protocol log/table should remain empty when no data on the input.

Actual behaviour

1)- In menu Analyze, Save CSV produce a message : "Not enough free space for this record" even if the storage has 16MB free space.

2)- In the Decoder interface, the F1 label is "Proto" where it should be "Protocol" (space is enough).

3)- In RS232 protocol, with no probe connected, there is a continuous flow of data (random).

DEC00007.CSV

SCR00003-35EA2143

Setup

No response

Activity

  1. added
    bug
    on Nov 26, 2026
  2. changed the title [-]Scope : Decoder Save CSV Issue[/-] [+]Decoder: Decoder Save CSV Issue[/+] on Nov 26, 2026
  3. bve commented on Nov 26, 2026

    @bve
    Owner

    Fixed in OpenOSC 0.21.19. Analyze → Save waveform CSV checks the exact file size when needed and shows the required and available space if the file cannot fit. Decoder → Save CSV now explains why a completed decode is unavailable, and the F1 label reads Protocol. RS-232 Auto hysteresis filters noise near the default 0 V threshold.

    I verified waveform CSV, I²C and SPI exports, and RS-232 decoding on a ZT703S.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment