Instruments

Instruments read the pulse at a location ("at": "end", "input", or an element id) and report what a real device would show, including its resolution. They never modify the pulse. Values go to summary.json -> instruments[], arrays to result.npz (prefixed with the instrument's name) and to <name>.csv. Example 17 (docs/examples/Diagnostics/instrument-tour) uses every kind once. Every option of every instrument is listed in the option reference.

Every instrument is a device model. Its parameters (resolution, range, sampling, acceptance, detector) are declared with units and defaults, and every output carries a settings block with the value of each parameter that was used -- given or default -- so a results file says exactly which instrument it describes:

"instruments": [ { "name": "spec", "kind": "spectrometer", "values": { "fwhmM": 1.1e-8, ... },
                   "settings": { "resolutionM": 2e-10, "points": 2048, "normalize": "peak", "rangeM": "default: the spectrum's support", ... } } ]

Several of the same. Instruments can be repeated anywhere, also several of the same kind at one place. Give each a name; a repeated or missing name is made unique (spectrometer_plate, spectrometer_plate_2, ...), with a warning when a name you gave is used twice.

kind parameters reports
energyMeter (powerMeter) rangeM [min, max], polarizerDeg, repRateHz energyJ, averagePowerW, transmissionFromInput; with rangeM the edge bins count with the fraction of the bin inside the band (error falls as the square of the bin width)
spectrometer resolutionM (Gaussian FWHM), rangeM, points, polarizerDeg, normalize (peak / area / none = J/m), db, wavelengthOffsetM spectrum vs wavelength; peak, centroid, FWHM, RMS and -10 dB widths
scope (photodiode) riseTimeS or bandwidthHz, points detected power vs time (impulse-response limited), FWHM, peak
autocorrelator type: intensity (default) or interferometric; rangeS, points, resolutionS (delay resolution), shgAcceptanceFwhmM (crystal bandwidth) trace; AC FWHM and deconvolved FWHM (Gaussian sqrt2, sech^2 1.543); IAC peak-to-background (8:1; the delay axis is symmetric with an odd number of points, so it contains d = 0, step lambda0/(16 c); if points forces a coarser step a note says the fringes are under-sampled; the background is the separated-pulse level 2 sum I^2)
frog type: shg (default), pg (polarisation gate) or sd (self-diffraction); points (default 256), spectralResolutionM, shgAcceptanceFwhmM trace [delay, frequency], axes, marginals; frogType (FROG geometries)
xfrog (sfgFrog) reference (the known gate pulse, a run-spec pulse object), points, spectralResolutionM cross-correlation FROG trace at the sum frequency; the reference field is stored with it (FROG geometries)
dscan material + insertionM [min, max], or gddRangeS2 [min, max]; steps, spectralResolutionM, shgAcceptanceFwhmM SHG trace vs insertion/GDD, integrated SHG, FWHM vs scan, optimum
contrast resolutionS, windowS, dynamicRangeDb intensity in dB; pre/post-pulse contrast; pedestal energy fraction
polarimeter points, extinctionRatio Stokes s1, s2, s3 (integrated and vs wavelength), degree of polarisation, azimuth, ellipticity
pulseSummary every scalar pulse metric (see results.md)
compressor optimizeTod (true), maxN ideal compressor: the GDD/TOD that maximises peak power, compressed FWHM, Strehl
conversion at: an active element (crystal) energy and photon number of the main output and every by-product, aux energy in, efficiencies (vs input, vs all energy in), by-product centroids
phaseMatchingCurve at: a crystal; scan: angle (rangeDeg), wavelength (rangeM), temperature (rangeK), period (rangeM); points low-conversion sinc²(ΔkL/2) and Δk vs the scan variable, peak, FWHM (Crystals)
delayScan at: a crystal with aux beams; fromS, toS, points re-runs the crystal on the pulse that entered it for each aux delay: output energy vs delay, optimum
beamProfile at: after an element with a transverse model fluence vs radius (radial) or x / y cuts and map (cartesian); second-moment radius about the centroid, M² (m2, and m2Y on 2-D grids), true peak fluence (Spatial models)
crossCorrelator (thirdOrderCorrelator) rangeS, stepS / points, fineStepS + fineRangeS, dynamicRangeDb (110), noise, resolutionS, shgAcceptanceFwhmM, polarization, probeDelaysS, gateArm third-order correlation vs delay (linear and dB, with and without noise), pre/post-pulse contrast and their delays, contrast at chosen delays (below)

Parameters every instrument shares

key what it does
at, name where the instrument sits, and its name in the results
sampling what part of the beam it sees: near / far field, pinhole, fibre, slit (below)
filter a spectral filter model in front of it -- optics, a band-pass, the detector's spectral response; applied to the field before it is measured (any filter model: constant, table, gaussian, expression, callback)
detector the read-out of the recorded trace (below)
axis where a curve instrument takes its readings: explicit points or a scan grid, pixel / bin width, resolution (below)
measured your recording, compared on its own axis, range and resolution (Measured data); used by fits

Every instrument output carries a settings block with each parameter in effect (given or default), so a results file states exactly which device -- resolution, range, sampling, detector -- produced each reading. Virtual instruments (pulseSummary, compressor, wigner, fieldProbe, spaceTime, beamProfile, conversion) are ideal by design: they report the field itself.

Detector model

{ "kind": "spectrometer", "resolutionM": 0.2e-9,
  "filter": { "kind": "table", "wavelengthsM": [900e-9, 1100e-9], "transmission": [0.6, 0.9] },      // grating + detector response
  "detector": { "calibration": 1.0, "noise": 0.01, "dynamicRangeDb": 40, "bitDepth": 12, "seed": 1 } }
detector key meaning
calibration reading / true value (e.g. 1.03 for a power meter reading 3 % high)
noise multiplicative noise, relative rms
noiseFloor or dynamicRangeDb additive noise rms relative to the trace peak
saturationLevel readings are clipped at this fraction of the trace peak (0: none)
response spectral response of the detector, {wavelengthsM, values} (e.g. the quantum efficiency): spectral instruments only; linear in wavelength, held at the ends
counting "photons": the energy spectrum is first converted to photon counts (x lambda / lambda at the centre of the axis); default "energy"
bitDepth ADC quantisation, full scale at the trace maximum
seed random seed: the same seed gives the same noise

It acts on the recorded arrays and readings of the spectrometer, scope, autocorrelator, FROG, d-scan and energy meter, and the width and peak readings (FWHM, centroid, peak) are then re-evaluated on the detected trace. Cameras (sensor) and the cross-correlator have detector models of their own. Leave the detector out when you fit a model to measured data: the recording has the noise, the model should not.

Detector presets

A detector you use again -- with its calibrated floor, saturation level, attenuation segments or quantum-efficiency table -- is defined once and named. The engine contains no instrument's numbers; they are yours (a manual, a calibration). In the run spec:

"detectors": {
  "my-xc": { "instrument": { "saturationLevel": 0.9, "noiseFloor": 1e-11,
                             "segments": [ { "fromS": -2e-12, "toS": 2e-12, "attenuation": 1e-3 } ] } },     // fills what the instrument does not set
  "my-ccd": { "detector": { "response": { "wavelengthsM": [900e-9, 1000e-9, 1100e-9], "values": [0.45, 0.35, 0.08] },
                            "counting": "photons", "noise": 0.005, "bitDepth": 16 } }
},
"instruments": [ { "kind": "crossCorrelator", "detector": "my-xc" },
                 { "kind": "spectrometer", "detector": "my-ccd" },
                 { "kind": "spectrometer", "detector": { "preset": "my-ccd", "noise": 0.02 } },           // a preset with overrides
                 { "kind": "spectrometer", "detector": { "preset": "presets/ccd.json" } } ]                 // or a JSON file (relative to the run spec)

A preset has an instrument object (keys filled into an instrument that does not set them: saturationLevel, segments, noiseFloor, dynamicRangeDb, resolutionM, ...; the instrument's own keys win) and a detector object (the read-out above; the instrument's keys given next to "preset" win). An unknown preset is a warning that names it. Python: build.detector_preset, build.detector, build.detector_response, run_spec(detectors=...).

Resolution and acceptance

  • Spectrometers (resolutionM, FROG and d-scan spectralResolutionM): Gaussian instrument function of the given FWHM in wavelength.
  • SHG crystals (shgAcceptanceFwhmM on autocorrelator, FROG, d-scan, cross-correlator): the phase-matching window of the doubling crystal, as a Gaussian of the given FWHM in SH wavelength, applied to the sum-frequency signal. In a FROG it narrows the frequency marginal; an intensity autocorrelation of a Gaussian pulse with linear chirp keeps its shape (only the signal drops) -- check the FROG or d-scan when the crystal may be too thick.
  • Delay (resolutionS on autocorrelator, contrast, cross-correlator): Gaussian smoothing along delay.
  • Detection bandwidth (scope riseTimeS / bandwidthHz).
  • Polarisers (polarimeter extinctionRatio, analyser polarizerDeg on spectrometers and energy meters).
  • Sampling of the axis (axis, every curve instrument): where readings are taken and what each integrates over (below).
  • Ranges and grids: spectrometer rangeM and points; FROG points; d-scan insertionM / gddRangeS2 and steps; cross-correlator rangeS, stepS, fineStepS, dynamicRangeDb; camera pixels, pixelSizeM, bitDepth; delay scan fromS, toS, points.

Axis: where readings are taken

A real instrument reads at particular points: the pixels of a spectrometer, the delay steps of an autocorrelator or cross-correlator, the samples of an oscilloscope. Each reading integrates over its pixel or bin. axis models this for every curve instrument: spectrometer, scope, autocorrelator, crossCorrelator, polarimeter, contrast, delayScan, beamProfile and phaseMatchingCurve.

{ "kind": "spectrometer", "resolutionM": 0.3e-9,
  "axis": { "values": [1020.1e-9, 1020.5e-9, ...], "binWidth": "pixel" } }      // your spectrometer's pixel wavelengths

{ "kind": "autocorrelator", "axis": { "from": -2e-12, "to": 2e-12, "step": 20e-15 } }   // a coarse delay scan

{ "kind": "scope", "axis": { "step": 50e-12, "binWidth": "pixel", "resolution": 200e-12 } }   // 20 GS/s sampling
key meaning
values explicit reading positions, in the axis's SI unit (m for wavelengths, s for delays and time)
from, to, step / points a regular grid (default: the simulated range and number of points)
binWidth what each reading integrates over: a number, or "pixel" (the spacing to the neighbours); default: point readings
resolution a Gaussian blur (FWHM) applied before binning, in addition to the instrument's own resolution keys

All of the instrument's curves are resampled together. dB curves are recomputed from their linear counterparts, so a cross-correlator keeps its noise floor and dynamic range. Readings outside the simulated range read zero (no light there), with a note. The width and peak readings (FWHM, centroid, peak) are re-evaluated on the sampled curve, and a detector then acts on the sampled readings. Recorded data (measured) are compared on their own axis in any case. axis matters when the sampling itself changes what is measured (coarse steps, wide pixels), and for exporting simulated data on your instrument's grid.

crossCorrelator

A third-order cross-correlator (Sequoia, Tundra, ...) measures the contrast of a pulse over 10+ orders of magnitude: the pulse is frequency doubled, the SH gate G(t) = |E²(t)|² is scanned across the fundamental, and the third harmonic is detected,

S(τ) = ∫ I(t) G(t - τ) dt.

{ "kind": "crossCorrelator", "name": "xc", "at": "end",
  "rangeS": [-200e-12, 100e-12], "stepS": 0.5e-12, "fineStepS": 20e-15, "fineRangeS": 3e-12,
  "dynamicRangeDb": 110, "probeDelaysS": [-50e-12, -10e-12] }
  • The gate is the pulse itself (the SH of the dominant polarisation), so the gate's own pedestal is folded in, as in the real device. A post-pulse at +T of relative intensity ε also appears as a weaker artefact (~ε²) at -T through the gate.
  • The detector floor is dynamicRangeDb below the peak (or noiseFloor), with a fluctuating background unless "noise": false; resolutionS adds the delay resolution of the scanner, shgAcceptanceFwhmM the finite SH bandwidth of the doubling crystal.
  • Delayed light. With grid.tail (Delayed light) the SHG gate is made from the main field only, and the probe channel (linear in intensity) from main + tail: a ghost or an echo appears as the real device sees it.
  • Resolution. With resolutionS the trace, the noise floor and the reported contrasts are all relative to the peak of the resolution-limited trace: smoothing smears a distant pre-pulse and the main peak alike, so a 1e-6 pre-pulse still reads -60 dB, and the floor stays dynamicRangeDb below the (smoothed) peak.
  • Elements in the gate arm only (gateArm). A real correlator makes the second-harmonic gate in its own arm, and an element there (a plate whose ghosts replicate the gate, a filter, dispersion) does not touch the fundamental. Give "gateArm": {"elements": [ ... ]} (any element list of a system file, optional "materials"): the second harmonic of the dominant polarisation, with the crystal acceptance applied, is sent through them compiled at lambda0/2, so glass and coatings are evaluated at the second-harmonic wavelength, and the gate is |E_SH|^2 of the result. The probe channel does not see them. The time window of the instrument is used: it must hold the replicas (grid.windowS, or grid.tail). A plate with ghosts gives replica k of relative level R^(2k) (R: the face reflectance at lambda0/2) at -2 k n_g d / c (the gate is delayed against the probe), as in the single-shot cross-correlator of Ma et al., High Power Laser Sci. Eng. 2018, doi 10.1017/hpl.2018.57 (Fig. 4c). Elements before the instrument in the system act on both arms. The run is linear: a nonlinear element in the gate arm sees an arbitrary power scale (the gate is scaled to the energy of the fundamental pulse).
  • Saturation and range stitching (generic, set by you -- no instrument's numbers are built in). saturationLevel clips the reading (signal x attenuation + floor) at that fraction of the unattenuated peak; segments is a list of {fromS, toS, attenuation (or gain), noiseFloor}: the signal in that delay range is measured through the attenuation (a power factor) and divided by it afterwards, so its floor rises as 1/attenuation and a strong peak no longer saturates. Example: "saturationLevel": 0.5, "segments": [{"fromS": -2e-12, "toS": 2e-12, "attenuation": 1e-3}]. Notes count the saturated points.
  • Reports: fwhmS, noiseFloor, dynamicRangeDb, prePulseContrastDb / prePulseDelayS and postPulseContrastDb / postPulseDelayS (strongest feature outside windowS), contrastAt_m50ps_db ... for each of probeDelaysS; arrays delay_s, signal, signal_db, signal_ideal.

The pulse needs its pedestal: give the input one (run spec: pedestal), let ghosts of plates and ND filters (components) and nonlinear conversion of post-pulses create pre-pulses, and widen the window when the features are far from the peak. Tutorial 5 follows a pre-pulse through an amplifier chain.

FROG geometries

frog type selects the gate (all with one x polarisation; delay tau, gate pulse delayed by tau):

type signal field carrier of the signal
shg E(t) E(t - tau) 2 w0
pg `E(t) E(t - tau)
sd (self-diffraction) E(t)^2 E*(t - tau) w0

SHG-FROG traces are symmetric in the delay and cannot tell E(t) from E*(-t); PG and SD traces can. For a linear chirp omega(t) = w0 + b t the intensity-weighted frequency centre of the row at delay tau is w0 + (2/3) b tau (PG) and w0 + (4/3) b tau (SD) (Kane and Trebino, Opt. Lett. 18, 823 (1993), appendix C); FrogKindsTests checks the engine's traces against these.

xfrog (alias sfgFrog) gates the pulse with a known reference R of any carrier (reference: the pulse object of a run spec, e.g. {"shape": "gaussian", "centerWavelengthM": 800e-9, "durationFwhmS": 30e-15, "chirp": {"gddS2": 195e-30}}): the signal E(t) R(t - tau) is at the sum frequency w0 + w_R. The reading stores the type (frogType) and, for XFROG, the reference field (reference_time_s, reference_field_re, reference_field_im, reference_omega_rad_s), so retrieval from it needs nothing else.

singleShotCorrelator

Time-to-space cross-correlator (Ma et al., High Power Laser Sci. Eng. 6, e61 (2018)): the pulse under test and a sampling pulse (its second harmonic, samplingPulse: "secondHarmonic", or the pulse itself, "fundamental", for an autocorrelator) cross at an angle in a crystal. The signal at position x across the overlap measures the correlation S(tau) = integral I_PT(t) I_S(t - tau) dt at tau = tau0 + x * slope, so a camera reads a delay trace in one shot.

key meaning
crossingAngleDeg or externalCrossingAngleDeg full angle between the beams inside the crystal, or in air (symmetric incidence, each beam refracted with its own index)
beamWidthM width W of the overlap across the signal; window T = W x slope
crystalLengthM L; geometric resolution dtau = L |(n_g,T - n_T sin^2 a) / cos a - (n_g,S - n_S sin^2 b) / cos b| / c
crystal (default BBO) or indexPt, indexSampling, groupIndexPt, groupIndexSampling ordinary (x) index of the library crystal, or given indices
delayIndex group (default) or phase: with phase the resolution is exactly Ma et al.'s eq. (2), L |n_T cos a - n_S cos b| / c
points, delayOffsetS, dynamicRangeDb, noise, seed camera pixels across W, delay at the centre, floor

The angles a, b of the two inputs to the signal follow from k_T + k_S = k_signal with the phase indices (k_T sin a = k_S sin b). The slope (n_T sin a + n_S sin b) / c is eq. (1): beams that enter from air at fixed angles keep their tangential wave number, so the phase index enters. Outputs: x_m, delay_s, signal, signal_db, signal_ideal, signal_raw; values windowS, resolutionGeometricS, resolutionPixelS, resolutionS, delayPerPositionSPerM, angles, windowCoverage, fwhmS, pre / post-pulse contrast.

What is simulated. The correlation of the temporal intensities at the instrument, a rectangle of width dtau (each position collects all depths of the crystal) and the pixel average, with a flat-top overlap. Area is conserved: the raw signal integrated over delay equals (integral I_PT dt)(integral I_S dt), windowCoverage reports how much of it falls inside the window. What is not. The beams are not propagated through the crystal here: no conversion efficiency or depletion, no beam profile across the overlap, no walk-off, no air scatter of the signal, no stripe filter, no reflections between the crystal faces. The engine's own spatial model (a Branched sfg crystal with two crossed arms on a Cartesian grid) does reproduce the mapping: the test SpatialSfgOfTwoCrossedBeams_PutsAReplicaAtTheInstrumentsPosition puts replicas at x = T c / (2 sin(10 deg)) at 20 degrees (slow, about 2.5 minutes).

Sampling: where the instrument looks

By default an instrument sees the whole beam: the spectrum, autocorrelation or power integrated over the full transverse profile. Real instruments often do not. A spectrometer might be fed by a fibre at the focus of a lens, a photodiode might sit behind an iris, and an autocorrelator might only accept the centre of the beam. When the pulse varies across the beam (SPM is stronger on axis, a Kerr lens, spatial chirp, a thermal lens), these all give different answers. Add "sampling" to any non-imaging instrument:

{ "kind": "spectrometer", "name": "fibre_spec", "at": "plate",
  "sampling": { "plane": "far", "focalLengthM": 0.1,
                "aperture": { "kind": "fibre", "radiusM": 25e-6, "na": 0.22 } } }
key default meaning
plane near near: the beam as it is at that point. far: the focal plane of an ideal lens of focalLengthM (the Fourier transform of each frequency's field).
focalLengthM 0.2 far field only
propagateM 0 near field: free-space propagation before the aperture (negative: back towards a waist)
aperture.kind pinhole if an aperture is given, else integrated integrated (whole beam), pinhole, iris (both circular), fibre (circular core, plus an NA check), slit (a vertical strip of widthM across the whole beam), point (the local spectrum at one point; shape only, no transmission)
aperture.radiusM 50 µm (fibre 25 µm) pinhole / iris / fibre core radius
aperture.widthM 2 × radius slit width
aperture.centerXM, centerYM 0 aperture position (off-axis sampling)
aperture.na 0.22 fibre: warns if the focused beam's convergence angle exceeds the NA

What is computed. For every frequency ω, the field E(x, y, ω) on the chosen plane is integrated over the aperture. This gives a power transmission T(ω) and an aperture-averaged spectral phase φ(ω). The instrument then measures the pulse with its spectrum multiplied by √T(ω)·e^{iφ(ω)}. So a spectrometer shows the local spectrum, and an autocorrelator or FROG shows the local pulse, including the spectral phase variation across the aperture.

  • The field comes from the carried transverse profile after an element with a transverse model. On a 1-D line, it is an ideal Gaussian from the instrument's "beam", the element's beam model or the run's input beam (the camera rules, see Camera). A 1-D Gaussian has the same spectrum everywhere in the near field, so sampling only matters there through the wavelength dependence of the far-field spot. That spot is smaller for shorter wavelengths, so a small fibre in the far field couples the blue side slightly better.
  • The far field uses the exact Fourier (cartesian) or Hankel (radial) integral of each frequency's field, scaled by λf, as the far-field camera does. It is normalised so that ∫|u|² dA equals the near-field power.
  • The instrument reports sampling (a description) and samplingTransmission (the fraction of the pulse energy that passes the aperture).
  • Imaging instruments (camera, wavefrontSensor, beamProfile, spaceTime) ignore sampling, because they already image a plane of their own (imaging).

Example 11 shows the effect clearly: after a self-focusing plate, a 15 µm pinhole on axis sees a 13 % wider spectrum and an 11 % shorter autocorrelation than the whole beam.

Recorded data (measured)

Any instrument can be given your recording ("measured": "data/spectrum.csv", a CSV, NPZ or JSON file). The run compares it with the simulated reading on the recording's axis, only where the simulated instrument has readings, after blurring the simulation to the recording's resolution when the recording is coarser. See Measured data. A "fit" in the run spec (fitting plugin) fits transform elements and other parameters against it, and instrument-response parameters (instrument:NAME/resolutionM) can be fitted too.

Camera

One instrument, camera, images the beam. imaging says what the sensor sees, sensor describes the detector:

{ "kind": "camera", "name": "focus", "at": "plate",
  "imaging": { "plane": "far", "focalLengthM": 0.5, "magnification": 1 },
  "sensor":  { "pixels": [256, 256], "pixelSizeM": 5.5e-6, "bitDepth": 12, "readNoiseCounts": 2,
               "backgroundCounts": 40, "qe": { "kind": "table", "wavelengthsM": [700e-9, 900e-9], "transmission": [0.6, 0.3] } } }
imaging key default meaning
plane near near: the beam at the instrument's location; far: the focal plane of an ideal lens of focalLengthM (exact Fourier / Hankel integral of every frequency's field, scaled by λf); relay: the near field after the relay steps
propagateM 0 near field: free-space propagation before imaging (negative: back to a waist, e.g. to image a focus)
focalLengthM 1 far field: focal length
magnification 1 imaging magnification onto the sensor (pixel size / magnification = pixel size in the object plane)
relay — list of thin lenses {"focalLengthM": f} and gaps {"gapM": d}, applied in order with exact diffraction on the field's grid (good for relays that do not change the beam size much; use magnification for large image scales)
sensor key default meaning
pixels 256 n or [nx, ny]
pixelSizeM auto pixel pitch on the sensor; auto fits ~5 beam radii across
centerXM, centerYM 0 sensor centre in the object plane
qe 1 quantum efficiency vs wavelength: any spectral-filter model (constant, table, gaussian, expression, "@callback"); the image weights each colour by it
bitDepth 0 (ideal) ADC bits; with bits the image is quantised and a raw counts array is added
fullScaleJPerM2 1.1 × peak fluence at full scale
readNoiseCounts 0 read noise (rms counts)
backgroundCounts 0 dark / background offset (in counts; removed again in the fluence image)
hotPixelFraction 0 fraction of stuck (saturated) pixels
noiseSeed 1 random seed

bandM [λmin, λmax] (top level) adds a band-pass filter. Sensor keys may also be written at the instrument's top level (as in format version 1). The kinds nearFieldCamera and farFieldCamera of version-1 files are migrated automatically to camera with imaging.plane near / far (Schemas & migrations).

Reports. Fluence image fluence_j_m2 [x][y] (J/m²) with axes x_m, y_m, and counts with a bit depth; centroid, 1/e² radii (D4σ/2) in x and y, ellipticity, peak fluence, energy in the frame, saturated-pixel fraction, imagingPlane; near field: M²; far field: divergence half-angles (radius / f), energy inside the 1/e² ellipse and the Strehl ratio (peak vs the same near-field amplitude with a flat wavefront).

Where the field comes from. After an element with a transverse model ("transverse": {"model": "radial" | "cartesian"}) the camera images the real simulated beam. On a 1-D element it images an ideal Gaussian from the instrument's "beam", the element's beam model or the run's input beam, and says so in its notes.

Wavefront sensor

kind parameters reports
wavefrontSensor pixels (lenslets, 64), wavelengthM (default: spectral centroid), threshold (1 % of peak fluence), apertureRadiusM (default the 1/e² radius), halfWidthM, zernikeTerms (15, max 36), propagateM, bandM wavefront map in waves (piston and tilt removed), slope maps slope_x_rad, slope_y_rad (λ ∂W/∂x and ∂y of the full wavefront, tilt kept, NaN outside the threshold, [x][y] on x_m, y_m: the zonal input of a control matrix), Zernike coefficients (Noll, rms-normalised, waves), rms / PV, rms without defocus, radius of curvature from defocus (R > 0 diverging; the equivalent in air: R in a medium / n), astigmatism, coma, spherical

Validated against Gaussian optics (DiagnosticsTests): near-field radius and refocused waist < 1 %, far-field radius λf/(πw0) < 0.5 %, wavefront curvature R(z) < 1 % (radial) / 2 % (cartesian), Zernike fit exact; PluginTests.Camera_UnifiedImaging_*: a 2f relay reproduces the far field (< 3 %), the old kinds match camera.

Virtual instruments

Virtual instruments read things no real device can measure directly -- the complete complex field, its absolute phase, both time and frequency at once, the field at every point of the beam vs time. They are flagged "virtual": true in their values and drawn with a square (□) marker and (virtual) titles in the plots, so their output is never mistaken for a lab measurement. (Ideal-device instruments such as pulseSummary, conversion, phaseMatchingCurve and delayScan are also shown as virtual.)

kind parameters reports
wigner points (256 time rows), gateFwhmS (> 0: spectrogram with a Gaussian gate), maxN Wigner distribution W(t, ω) [time][frequency] (normalised to max
fieldProbe threshold (1e-3), realField (false), maxN temporal phase and instantaneous wavelength vs time, group delay and GDD vs wavelength, carrier-envelope phase, linear chirp rate; with realField the real field Re[A(t) e^(-iω0t)] on the optical cycle
spaceTime points (512), maxRadiusM (2.5 w), axis (x/y, cartesian) intensity I(r, t) and spectral fluence S(r, λ) across the beam, arrival time, local FWHM and centroid wavelength vs position; pulse-front delay and spatial chirp at r = w, on-axis vs r = w duration

Where an instrument sits

ultrafast-plot numbers the instruments in system order (input first, then element by element, end last), draws them on the system schematic (system.png, and the strip at the top of overview.png) and prefixes every panel title with the number and location, e.g. ③ photodiode stretched · after probe_stretched. Put instruments wherever you would put them on the table -- "at": "input", an element id (after that element), "end" -- or add a probe element as a named place in the beam (see docs/examples/LaserSystems/stretcher-compressor/run.json).

Conventions: Stokes s3 = +1 for the "rcp" input polarisation; FROG/d-scan use the x polarisation.

Add your own by implementing IInstrument and calling InstrumentFactory.Register("myKind", spec => new MyInstrument()) (Extending).